Liquid air separation plant temperature measurement instrument common failures

Liquid air separation plant temperature measurement instrument common failures
Jun 10, 2022
Author: DEAR AST

Thermocouple thermometer common fault for the indication of low value, or stop at zero. This is due to the thermocouple generated problems, may be poor contact or broken circuit, generally do not appear to indicate the phenomenon of high.
In order to ensure the measurement accuracy, the use of the cold end of the temperature compensation should be correct (commonly used compensation wire method), and pay attention to the thermal electrode and the earth between a good insulation.
RTD thermometer common fault: sometimes the pointer always points to the maximum. This is a resistance thermometer temperature measurement element break, so that the resistance of infinity, or poor contact between the two ends, contact resistance; another phenomenon is the pointer pointing to zero, or deflected to the negative direction. The reason is a short circuit in the temperature measuring element, resistance to zero, or due to the resistance thermometer protection casing leakage or icing caused by; there is another phenomenon is the pointer does not move, which is due to the bridge itself out of the problem, not working caused by.
In use to strictly prevent moisture into the RTD protection tube, after the display meter must not be open or short-circuit. In the inspection of the RTD, the measurement current should not be greater than 5mA.

Translated with www.DeepL.com/Translator (free version)

DEAR Air Separation Solve YourLiquid air separation plant temperature measurement instrument common failuresYou can also learn about otherLiquefied air separation FAQ about DEAR Air Separation, and solve the problem for you.
Didn’t find the problem encountered?
Please contact live chat or email for quick help.
Related FAQ
More Related FAQ
Problems in the process of pressure swing adsorption of Biogas
2026.08.07
The main measures taken for the design of oxygen piping in small air separation unit. 1、The piping material of the high-pressure oxygen gas pipeline is made of stainless steel to reduce the possibility of iron oxidation and combustion under intense working conditions. 2、Partial commercial flow rate piping using copper brocade alloy (Monel); 3, the use of different conditions of the pipeline proposed different degreasing and cleanliness requirements in the selection of Tong expansion joint for...
2026.08.06
Need more information?
Contact us for more details about our air separation equipment, solutions and services.
Products
  • Large air separation
  • Small air separation
  • High purity nitrogen air separation
  • Liquefaction air separation
  • Complete air separation
  • Biogas purification
  • Air separation parts
Need more information?
Tel:+86-19837850537
© Kaifeng DEAR Air Separation Industry Co., Ltd.. All rights reserved.
Online
Inquiry
Email
Top