Liquid air separation plant temperature measurement instrument common failures

Liquid air separation plant temperature measurement instrument common failures
Jun 10, 2022
Author: DEAR AST

Thermocouple thermometer common fault for the indication of low value, or stop at zero. This is due to the thermocouple generated problems, may be poor contact or broken circuit, generally do not appear to indicate the phenomenon of high.
In order to ensure the measurement accuracy, the use of the cold end of the temperature compensation should be correct (commonly used compensation wire method), and pay attention to the thermal electrode and the earth between a good insulation.
RTD thermometer common fault: sometimes the pointer always points to the maximum. This is a resistance thermometer temperature measurement element break, so that the resistance of infinity, or poor contact between the two ends, contact resistance; another phenomenon is the pointer pointing to zero, or deflected to the negative direction. The reason is a short circuit in the temperature measuring element, resistance to zero, or due to the resistance thermometer protection casing leakage or icing caused by; there is another phenomenon is the pointer does not move, which is due to the bridge itself out of the problem, not working caused by.
In use to strictly prevent moisture into the RTD protection tube, after the display meter must not be open or short-circuit. In the inspection of the RTD, the measurement current should not be greater than 5mA.

Translated with www.DeepL.com/Translator (free version)

DEAR Air Separation Solve YourLiquid air separation plant temperature measurement instrument common failuresYou can also learn about otherLiquefied air separation FAQ about DEAR Air Separation, and solve the problem for you.
Didn’t find the problem encountered?
Please contact live chat or email for quick help.
Related FAQ
More Related FAQ
What safety problems should be paid attention to when contacting nitrogen
2025.10.17
There is an open heat pump argon extraction process.   The new single column process can only use the open heat pump argon extraction process. Taking 10000cm³ air separation as an example, the main parameters are described as follows.   1、 The quantity of argon fraction is 4200-4300cm3 / h, the content of argon is 10%, the content of nitrogen is 1%, and the rest is oxygen.   2、 The theoretical plate number of the whole column is about 100.   3、 Two heat pump dist...
2025.10.16
Need more information?
Contact us for more details about our air separation equipment, solutions and services.
Products
  • Large air separation
  • Small air separation
  • High purity nitrogen air separation
  • Liquefaction air separation
  • Complete air separation
  • Biogas purification
  • Air separation parts
Need more information?
Tel:+86-19837850537
© Kaifeng DEAR Air Separation Industry Co., Ltd.. All rights reserved.
Online
Inquiry
Email
Top