当前位置: Home / Faq / Large air separation FAQ /

Relationship between failure of centrifugal liquid oxygen pump and seal pressure regulation

Relationship between failure of centrifugal liquid oxygen pump and seal pressure regulation
Jul 22, 2025
Author: DEAR AST
Question Reply

Due to improper adjustment of seal gas pressure during operation, liquid oxygen pump can not pump liquid or leak.

 

From the quality point of view, when the sealing pressure is too high, there will be gas leakage into the pump through the labyrinth seal, resulting in the impeller with gas or even idling, resulting in liquid injection or pressure drop; when the air pressure before sealing is too low, there will be liquid oxygen leakage.

 

From the quantity point of view, when (p1-p3) > 0.005 ~ 0.01Mpa, the sealing gas will enter the pump shell and air will appear. When p3 > p1, it will leak.

 

The purpose of using seal gas (dry nitrogen) is to prevent or reduce the leakage of liquid oxygen, but it is not allowed to carry air. Therefore, the principle of adjusting the sealing pressure is to allow the pump to operate with a small amount of liquid oxygen leakage and vaporization. The pressure level is related to many factors, such as the sealing structure, the position of the vent hole, the clearance of the pump, etc. It is difficult to set data strictly, so it is necessary to explore in practice to find a suitable method.

 

In general, the pressure difference (p1-p3) before and after the seal gas is preferably in the range of 0.005 to 0.01Mpa. For example, some factories initially stipulated that the pressure of seal gas should be higher than the discharge pressure of liquid oxygen pump, but it is not suitable in practice, which usually leads to pressure drop. After that, when the seal gas pressure is reduced to 0.02MPa lower than the pump outlet pressure, the normal operation can be ensured.

DEAR Air Separation Solve YourRelationship between failure of centrifugal liquid oxygen pump and seal pressure regulationYou can also learn about otherLarge air separation FAQ about DEAR Air Separation, and solve the problem for you.
Didn’t find the problem encountered?
Please contact live chat or email for quick help.
Related FAQ
More Related FAQ
Why do molecular sieves in small air separation unit use bunk beds? Molecular sieve purifiers are used in both small and large air separation plants to purify the air. carbon dioxide, acetylene, and various hydrocarbons. Some molecular sieves small air separation unit use a double bed because the adsorption capacity of activated alumina is relatively large for air with high water content, but as the water content of the air decreases, the adsorption capacity decreases rapidly. The molecular s...
2026.01.29
It should contact with the front and rear processes when the piston compressor is ready to stop, we should pay attention to the following sequence: first, cut off the gas supply slowly. In order to prevent the outlet pressure from rising after the gas supply is cut off, open the vent valve at the same time. For the oxygen compressor, the vent valve should be opened slowly, because excessive opening may produce static electricity and cause fire. Then, stop the motor. After the motor stops, turn off the circulating oil pump, grease injector motor and cooling water valve. However, there are also exception cases. For example, in case of emergency shutdown, in order to save time and prevent the accident from expanding, the power supply can be cut off immediately to stop the compressor with load.
2026.01.28
Need more information?
Contact us for more details about our air separation equipment, solutions and services.
Products
  • Large air separation
  • Small air separation
  • High purity nitrogen air separation
  • Liquefaction air separation
  • Complete air separation
  • Biogas purification
  • Air separation parts
Need more information?
Tel:+86-19837850537
© Kaifeng DEAR Air Separation Industry Co., Ltd.. All rights reserved.
Online
Inquiry
Email
Top